How to set up your GPS Logger using the INNOTrack Platform
Help Guide: How to set up your GPS Logger using the INNOTrack Platform.
1. Log in to INNOTrack Plaform
- Go to: INNOTrack Login
- Enter your account details
2. Add Device
- Navigate to Devices β Add Device.
- Fill in the required information.
- ClickΒ OKΒ once complete.
3. Create Cargo Profile (Optional but Recommended)
- Go to Template Manage β Cargo Profile β Create Template.
- Enter the parameters (temperature, humidity, shock, light, etc).
- Save the profile. This allows quick set-up for multiple devices in the future.
4. Set Up the Device
- Go to Device Settings β Set up Device.
- Select the device to start (click on the β¦ next to the device).
- Select device for the shipment.
- Enter the required information relating to the shipment.
- Set Location Method to Cell and Wi-fi, GPS is not currently supported.
- Under Forbid Manual Stop, set to Yes if you want to prevent the logger from being stopped manually. This ensures the device can only be stopped via the platform software, maintaining complete data control.
- Enter Next
5. Configure Cargo Profile & Alarms
- Either select a pre-set cargo profile or manually configure thresholds.
- Set alarms for Temperature, Humidity, Light, and Shock depending on model of the logger.
- Enter email addresses to receive alarms (add additional recipients via Share information).
- Click Next.
6. Confirm Device Set-Up
- Review the setup on the confirmation screen.
- If everything is correct, click Confirm Set Up.
7. Start Logger
- Go to Overview β check status (should display Idle).
- On the physical device: the letter R will appear on the LCD within 1 hour β device is logging.
- On the platform: the logger status will update to In Transit.
IMPORTANT:Β If the logger is being set up via software, the operator must not press the physical buttons on the device. Using the buttons during software setup may stop or interrupt logging.
8. End Logging
Option A: Stop on Platform
- Go to Devices.
- Click the green In Transit icon for the device.
- On the right-hand side, click End.
Option B: Stop Manually
- On the device, press and hold Stop button for 5 seconds (only applicable if Forbid Manual Stop was not activated.
- If Forbid Manual Stop was activated logger needs to be connected to PC via USB cable.
- Once the letter R disappears, logging has stopped.
9. Reports
- After stopping the device, allow up to 1 hour for the logger to upload its final data.
- Once available, go to the Reports section and download the PDF report.
- Click on Load Data to refresh and retrieve the latest report.
- Select the logger by clicking the Status icon or the Eye icon on the same line.
- The results will display automatically, including:
- Recorded temperature/humidity/light/shock values (depending on setup).
- Alarm status (any thresholds exceeded).
Below you can see an example of how the INNOTrack platform can help you to monitor shipments in real time and access valuable data.
There are three different styles of GPS Logger in the Klipspringer range:
- Single-Use Temperature and Light Data Logger
- Multi-Use Temperature, Humidity, Location, Shock, and Light Data Logger
- Single-Use Temperature and Light Data Logger with an External Probe
To learn more about the options available and their different applications, click the button below.
You can also reach out to our support team on 01473 461800Β orΒ sales@klipspringer.com. Alternatively, you can share your details using the contact form below.Β
If you want to learn more about GPS Loggers or the INNOTrack Platform, our in-house team would be happy to help.
Common Electrode Misunderstandings: White Crystals and Clear Liquid in the Protective Cap
White crystals on the tip of your electrode? Clear solution inside the protective cap?
We explain why this isn't cause for concern when unpacking your equipment.
Two of the most common queries relating to Electrochemical Analysis are:
- There are white crystals around my electrode, is the equipment still suitable for use or has something gone wrong?
- When I opened up my equipment, there was a clear solution in the protective cap that covers the electrode, is there a problem with the instrument?
In both instances, there is no cause for concern. To understand more, take a look at the video guide below. We have also put together a brief written overview if you prefer.
Video Guide
White Crystals on your Electrode

When opening up your new equipment, you may notice that white crystals have formed around your electrode. This is not a fault.
The protective cap placed over your electrode contains Storage Solution. This helps to protect the electrode and extend its lifespan.
When the Storage Solution dries, it is normal for white crystals (aka salt deposits) to form.
To prevent contamination, make sure to remove the white crystals before use. You can do this by using warm water and white roll to gently wipe them away.
Solution in your Electrode Cap

As mentioned above, your electrode will arrive at your site with a protective cap that contains Storage Solution. This solution helps to keep the electrode hydrated – ensuring its accuracy and extending its lifespan.
In some instances, sites have confused this liquid with evidence of a leak. However, it is just the Storage Solution doing its job.
To make sure the Storage Solution is kept inside the protective cap and doesn’t spill, we recommend unscrewing the cap with the electrode pointing down. The solution should be settled in the bottom of the cap.
So there you have it, two of the most common queries relating to electrodes – resolved!
Whether it’s white crystals around your electrode or clear solution inside the protective cap, this doesn’t have to mean a fault with your equipment. Instead, the most likely explanation is that the Storage Solution inside the cap is keeping your electrode safe and accurate.
If you have any questions about the information shared throughout this guide, you can reach out to our support team on 01473 461800Β orΒ sales@klipspringer.com. Alternatively, you can share your details using the contact form below.Β
If you want to learn more about our wide range of instrumentation, the Klipspringer team would be happy to help. Share your details below to arrange a free consultation.
How to Use Colour Coded Gloves in Food Production
Help Guide: How to Use Colour Coded Gloves in Food Production.
Colour coded gloves play a critical role in helping food manufacturing sites to maintain standards of safety, hygiene, and compliance.
The following guide outlines why they matter and which colours to use for different factory zones. It will also talk you through the process of establishing and maintaining your policy.
Alternatively, to view the different colour options available in our Pearl Range of Food-Safe Nitrile Gloves, click the button below.
Why Use Colour Coded Gloves in Food Production?
-
Reduces cross-contamination risk
Colour coded gloves help prevent the transfer of allergens, pathogens, and other contaminants between zones. This is crucial for maintaining high food safety standards. -
Improves employee awareness
Acts as a clear visual reminder of your site’s segregation policy, encouraging staff to follow correct hygiene practices consistently. -
Promotes accountability and error detection
Enables supervisors to spot non-compliance or glove mix-ups at a glance – reducing the risk of food safety incidents. -
Simplifies training and onboarding
The use of colour coded gloves is easy to understand, so will help new team members quickly learn and follow site protocols, reducing induction time and the cost of training. -
Strengthens audit readiness
Supports HACCP and BRCGS compliance by demonstrating clear, proactive contamination control and traceability measures. -
Integrates with wider colour coding systems
Colour coded gloves work seamlessly with other coded items – such as equipment, utensils, and shadow boards – to ensure a fully aligned strategy.
Which Colour Coded Gloves Are Best for Different Food Factory Zones?
There is no universal standard for colour coded gloves in food production. Instead, most food factories establish their own approach based on internal risk assessments and existing colour coding policies. That said, there are some common industry practices to consider. The examples below may serve as a useful starting point when developing or reviewing your own policy.
Blue Gloves – General Food Handling

Blue food safety gloves are typically used for general food factory applications. This includes the handling of raw fish, cooked meat, and ready-to-eat products.
Blue is widely recognised as a suitable choice for PPE because very few food products are this colour. Because of this, if a blue glove (or fragment) finds its way into a product through misuse or operator error, it will be much easier to find – a win for food safety, operational efficiency, and the reduction of food waste.
Red Gloves – Raw Meat Handling

Red food safety gloves are commonly used in areas where raw meat is inspected, handled, or prepared.
A key reason for this is that they are easily distinguishable from blue safety gloves which are typically used for general food factory applications, including the handling of ready-to-eat products. This helps to reduce the risk of cross-contamination, making it easier for teams to monitor compliance and identify any instances of glove misuse across zones.
Yellow Gloves – Cooked Meat Handling

Once meat is cooked, it is unlikely to be subjected to any further heat treatment. So, if it’s contaminated, there won’t be an additional step to kill bacteria before it’s eaten. This makes the prevention of cross-contamination between raw meat and cooked meat all the more important, as raw meat often contains harmful bacteria like Salmonella, E. coli, and Listeria. To reinforce this need for segregation, sites will often use yellow food safety gloves for the handling and processing of cooked meats.
Orange/Purple/Lime Gloves – Allergens
If a site has dedicated allergen and allergen-free zones, colour coding will play an important role in reinforcing segregation and reducing the risk of cross-contamination. Typically, it is the more unusual colours that are chosen for use with allergens.
Orange gloves and purple gloves are two popular examples, as they are unlikely to already feature in a site’s colour-coding policy, so could be the perfect choice if a new allergen-free line has just been added. Lime green is another useful option.
How Many Glove Colours Are Available for Food Factory Use?
Although we have highlighted some of the key colours above, there are plenty more to choose from. The Pearl Range of Single-Use Nitrile Gloves includes 16 colour options. As well as supporting your site’s colour coding policy, these gloves also offer the following benefits:
- Nitrile formulation for an improved fit
- Textured palms and fingertips for better grip
- Beaded cuff prevents liquid roll back
- Touchscreen compatible
- Medical-grade compliance
- Reduced risk of latex allergies (Type I)
















Establishing Your Colour Coded Glove Policy
-
Consistency is key!
As mentioned above, sites will typically use gloves that reflect their existing colour coding policy. This helps to avoid confusion, as mistakes are easier to identify and resolve. -
Don’t overcomplicate things
Only introduce additional colours when absolutely necessary. You want to ensure your policy is easy to follow and has the potential to grow as new products and applications are introduced. -
Speak to your supplier
If you have any questions or concerns relating to colour coding or the specific features of your gloves, speak to your supplier. -
Manage the risks
Carry out a thorough risk assessment and make sure your policy doesn’t overlook any obvious issues e.g. introducing white gloves to an area of your site that handles rice.
Maintaining Your Colour Coded Glove Policy
-
Clear and consistent records
Auditors will expect to see documentation of your colour coding policy and the thought process behind it. Having a central record will also help to drive standards of consistency across your site. -
Ongoing training and employee engagement
Although colour coding is easy to understand, make sure to explain the ‘Why?’ behind your policy. Effective segregation protects the future of your business and helps to save lives! -
Visual Reminders
Place signs or visual management boards at the entry points to each zone. You could also add reminders to any dedicated PPE stations or the bespoke mirrors in your changing rooms. -
Auditing and accountability
Colour coding and segregation should already be a key focus of your Internal Audits – add a note for your inspectors to check that your colour coded glove policy is being upheld.
That brings us to the end of our Help Guide on Colour Coded Gloves in Food Manufacturing. If you would like to learn more about our range of Pearl Gloves, click the button below.
Alternatively, if you have any questions about the information shared throughout this guide, you can reach out to our support team on 01473 461800Β orΒ sales@klipspringer.com. You can also share your details using the contact form below and one of our friendly team members will be in touch.Β
If you want to learn more about our wide range of PPE, the Klipspringer team would be happy to help. Share your details below and we’ll be in touch.
How to use and calibrate your Conductivity Pocket Tester
A step-by-step guide to calibrating your Klipspringer Conductivity Pocket Testers. Plus, everything you need to know about taking accurate measurements.
Conductivity can affect the taste, texture, and shelf life of a food or drink product. That is why teams across the industry are using Conductivity meters to ensure their output is safe and consistent.
To provide support in this area, we have developed a help guide for working with our own range of Conductivity Pocket Testers.
Featuring your choice of video and written instructions, it will take you through the process of calibrating, then using these pocket-size devices.
Calibrating your Conductivity Pocket Tester
It is important that you calibrate your Conductivity Pocket Tester before each use. You will need to calibrate it against the relevant Standard Solutions, for most sites this will be 1413 and 12.88.
1. Your Conductivity Pocket Tester is supplied with a cap containing Storage Solution to protect the probe and extend its lifespan. Remove this cap.
2. Press the on/off button to turn on your Conductivity Pocket Tester.
3. Pour the 1413 Standard Solution into a beaker. Place the probe directly into the Standard Solution and hold steady, no stirring is needed.
4. Press the calibrate button and wait for the device to stabilise. You will know the device has stabilised when a smiley face appears on the screen.
5. Press the βCALβ button and wait for the device to sync.Β
6. Remove the probe from the Standard Solution, rinse with distilled water, then gently pat dry.
7. Pour the 12.88 Standard Solution into a beaker. Place the probe directly into the second solution and repeat steps 4-6.Β
NOTE: Instead of pouring your Standard Solutions into beakers, another option is to use our innovativeΒ Easy-Dose Bottles. This will allow you to save on time and clean-up.Β
Using your Conductivity Pocket Tester
Before using your Conductivity Pocket Tester, you will need to calibrate the device. To do this, follow the calibration steps listed above.

1. Place the probe directly into your sample and hold steady, no stirring is needed.

2. The measurement is complete once the reading stabilises on the screen.
NOTE: To avoid damaging the components of your Conductivity Pocket Tester, only allow the tip of the device to touch the sample.

3. Remove the probe from the sample, rinse with distilled water, then gently pat dry.

4. Replace the protective cap and return your Pocket Tester to its case for safe storage.
Our Electrochemical Analysis range features pocket testers for pH, ORP, and Conductivity. There are also multi-parameter testers available.
If you have any questions about the products mentioned throughout this guide, you can reach out to our support team on 01473 461800Β orΒ sales@klipspringer.com. Alternatively, you can share your details using the contact form below.Β
If you want to learn more about our products and solutions, the Klipspringer team would be happy to help. Share your details below to arrange a consultation.
How to use and calibrate your pH Pocket Tester
How to calibrate your Klipspringer pH Pocket Testers, plus a step-by-step guide to taking accurate measurements.
Across the food and beverage production industry, teams are using pH meters to drive standards of safety, quality, and compliance.
To provide support in this area, we have developed a help guide for working with our own range of pH Pocket Testers.
Featuring your choice of video and written instructions, it will take you through the process of calibrating, then using these pocket-size devices.
Calibrating your pH Pocket Tester
It is important that you calibrate your pH Pocket Tester before each use. You will need to calibrate it against all relevant calibration points, for most sites this will be pH 4 and 7.
1. Your pH Pocket Tester is supplied with a cap containing Storage Solution to protect the probe and extend its lifespan. Remove this cap.
2. Press the on/off button to turn on your pH Pocket Tester.
3. Use the βCALβ button to switch the device to calibration mode.
4. Place the probe directly into the first Buffer Solution and hold steady, no stirring is needed. If a smiley face appears on the screen, this indicates the reading has stabilised.
5. Press the βCALβ button and wait for the reading to stabilise again.
6. Remove the probe from the Buffer Solution, rinse with distilled water, then gently pat dry.
7. Place the probe directly into the second Buffer Solution and repeat steps 4-6.Β
8. Press the βMODEβ button to change the device to measurement mode.
Using your pH Pocket Tester
Before using your pH Pocket Tester, you will need to calibrate the device and change the setting to measurement mode. To do this, follow the calibration steps listed above.

1. Place the probe directly into your sample and hold steady, no stirring is needed.

2. The measurement is complete once the reading stabilises on the screen.
NOTE: To avoid damaging the components of your pH Pocket Tester, only allow the tip of the device to touch the sample.

3. Remove the probe from the sample, rinse with distilled water, then gently pat dry.

4. Replace the protective cap and return your pH Pocket Tester to its case for safe storage.
Our Electrochemical Analysis range features pocket testers for pH, ORP, and Conductivity. There are also multi-parameter testers available.
If you have any questions about the products mentioned throughout this guide, you can reach out to our support team on 01473 461800Β orΒ sales@klipspringer.com. Alternatively, you can share your details using the contact form below.Β
If you want to learn more about our products and solutions, the Klipspringer team would be happy to help. Share your details below to arrange a consultation.
How to assemble your SegriScreen MK3
How to assemble your SegriScreen, plus everything you need to know about cleaning, maintenance, and storage
SegriScreens support effective segregation within the food production industry. Suited to everything from allergen segregation to washdown protection, they offer the following benefits:
- Mobile screens connect to form a continuous, connected barrier.
- Modular system - simply add or remove as many SegriScreens as required.
- Screens nest for convenient storage and transportation.
- Favoured by industry leads such as Bakkavor, Kerry, and Greencore.
- Hygienically designed for food industry applications.
- Add bespoke print for personalised branding, safety or instruction messages.
NEW! SegriScreen Design
Now with a new and improved design, Klipspringer SegriScreens also promise:
- Stronger design: bolted corner plates have been replaced with welds.
- New popper system: eliminates the need for rings and eyelets.
- 40% fewer components: reducing the risk of foreign body incidents.
- Easy assembly:Β one person can assemble in under 10 minutes.
As mentioned above, the new SegriScreen design is easier than ever to assemble, taking one person under 10 minutes. Even, so we wanted to produce an online Help Guide that you can return to if you ever need to remind yourself of the process.
SegriScreen Assembly – Video Instructions
SegriScreen Assembly – Written Instructions

1. Select a suitable flat, open area for assembly (minimum 3x3m). Lay out as shown, ensuring the male poppers on PART A and PARTΒ B face upwards. The pins on PARTΒ C and the loops on PARTΒ D should be pointing down.Β Β

2. FitΒ PART C (vertical left bar) and PART D (vertical right bar) into PART BΒ (horizontal bottom bar). To ensure there is a tight fit, it’s important that the components and fixing holes align.Β

3. Use one PART F (M6 x 20mm bolt) and one PART GΒ (M6 anti-vibration washer) on each side to fix the vertical bars (PARTS C and D) to the bottom horizontal bar (PART B). The suggested bolt torque value is 10 Nm.Β

4. Use oneΒ PART FΒ (M6 x 20mm bolt) and oneΒ PART GΒ (M6 anti-vibration washer) on each side to fix the vertical barsΒ (PARTS C and D) to the top horizontal bar (PART A). The suggested bolt torque value is 10 Nm.Β
NOTE: The M6 washers are dome shaped. They should be fitted so the larger diameter is in contact with the frame and the smaller diameter is in contact with the bolt face.

5. Lift the completed frame to an upright position and insertΒ PART EΒ (wheel bases) into each side. Use two PART F (M6 x 20mm bolts) and two PART GΒ (M6 anti-vibration washers) on each side to fix in place.Β
NOTE: The flat side of the wheel section should face outwards. Master screens have two sets of wheels. Additional/Extension screens have only one set of wheels.Β
Connecting your SegriScreens

SegriScreens can operate as a modular system, with screens added or removed as/when necessary. This will typically occur when the protection/segregation of a larger area of your factory is required.
To join multiple SegriScreens together, ensure the Additional/Extension screens only have one set of wheels.
Next, join the SegriScreens together by connecting the two male/female fittings on each vertical side. There are two male/female fittings on each side.
How to clean, maintain, and store your SegriScreens
When it comes to introducing any new equipment to your site, it is vital that you and your team are aware of the best approach to cleaning, storage, and maintenance.
This will help you to drive standards of hygiene and food safety. It will also help you to extend the lifespan of your equipment. With this in mind, we have put together a guide to effective SegriScreen maintenance.
Alternatively, if you are looking for a couple of quick pointers, the most important tips to remember are listed below.

1. SegriScreen bolts should be checked at regular three month intervals (or as specified by the policies at your food production site).
2. General cleaning and disinfection should be performed as per company procedure, factoring in the following guidelines:
- Normal detergent with warm water may be used for general cleaning.
- When cleaning your SegriScreens, the use of strong chemicals should be avoided.
- Ensure all chemicals are rinsed off your SegriScreens at the end of each clean.

Removing the SegriScreen PVC Panel
1a. Insert a screwdriver between the male and female snap fittings, then prise open.
1b. At the same time, pull the PVC screen panel away from the stainless steel frame.
Disposing of your SegriScreens
With the correct approach to cleaning, storing, and maintenance, your SegriScreens should stand the test of time.
However, if an unexpected incident should occur, replacement screens are available – allowing you to replace the material whilst making use of the existing base.
Alternatively, if you need to dispose of both the frame and the screen, you will need to separate the PVC screen panel from the stainless steel frame. You can then dispose of each part as per your site’s existing policies. This equipment is fully recyclable.
To learn more about SegriScreens or the other segregation solutions available, you can navigate the buttons below.
You can also reach out to our support team on 01473 461800Β orΒ sales@klipspringer.com. Alternatively, you can share your details using the contact form below.Β
If you want to learn more about SegriScreens or other segregation solutions, the Klipspringer team would be happy to help. Share your details below to arrange a free consultation.
How to Measure Equipment for a SegriCover
A guide to measuring your equipment for a SegriCover, exploring every aspect of the process, along with the key points to remember when taking your measurements.
Used by food production sites to maximise hygiene and minimise downtime, SegriCovers provide
a simple yet effective solution to equipment protection.
Instead of risking expensive control panels, machines, and production lines during washdowns, you can simply cover your equipment with a heavy-duty, waterproof shield that eliminates the need for unsafe and unsustainable single-use plastic.
With SegriCovers growing in popularity, we felt it was important to answer the most common questions surrounding this solution, and what better place to start than with a step-by-step guide to measuring your equipment?
NOTE: Before we get started, we wanted to highlight a key point to remember. Photos, especially annotated photos, are extremely helpful when it comes to sharing your measurements. This is also true of machine drawings and PDFs. If you are able to submit these resources, alongside your measurements, this will simplify the process of completing your SegriCover project.
How do you measure equipment for a SegriCover?Β
What should you be measuring?
The simplest approach is to measure the equipment itself, as opposed to trying to guess the measurements of the cover. We will then use the details of your machine to design a SegriCover that is the correct size.
The cardboard box technique
When it comes to measuring your equipment,
a helpful trick is to imagine you are measuring for a cardboard box that needs to sit neatly over the machine. To do this, you will need to measure the maximum allowed height, width, and depth. Β
You will also need to identify any obstructions that could prevent this imaginary box from covering your equipment. The most common examples are awkwardly-shaped machine parts and mounting points.

How do you account for machine parts and mounting points?
If your equipment has additional parts that extend beyond the central shape (e.g. pipes, wires, and light beacons), one option is to ensure the imaginary box is big enough to hold them, too. You will just need to be careful that this doesn’t disrupt the overall shape of your SegriCover and the equipment underneath is protected from damage. If you are concerned, the other option is to include cut-out requests in your measurements.Β
As for the mounting points, if they would be easy to cover, one idea is to factor them into your measurements. This could be an option if your equipment is attached to a free-standing bracket. However, if it would be impossible to pull a cover over the mounting points, for example, if your equipment has been bolted to the factory wall, you will need to request cut-outs.Β
How do you take measurements for a SegriCover withΒ cut-outs or clear windows? Β

Measuring for cut-outs
Cut-outs are a good option to explore if you are dealing with extremely large or awkwardly-shaped machine parts. This includes parts that donβt need to be protected during washdowns e.g. conveyor belts. It also includes parts that need to be protected from water ingress e.g. a large light beacon.
In the first instance, we would introduce the necessary number of cut-outs to ensure the relevant part(s) could pass through the SegriCover, instead of sitting inside and disrupting its shape. In the second instance, we would produce separate SegriCovers: one for the main piece of equipment and one (or more) for the extended part(s).Β
In both cases, you will need to take detailed measurements of the equipment part(s) that need to reach through the SegriCover: measuring how close the part(s) sit to the different sides and nearest corner of the machine, plus how tall and wide the part(s) are. Using these measurements, we will be able to introduce cut-outs that are the right size.
What fastenings will be used to complete this process?
If your SegriCover is going to wrap around machine parts instead of fitting over your equipment in a simple box shape, any cut-outs will need to be accompanied by a corresponding slit. This will allow your operatives to pull the cover on and off.
As for holding these slits in place, there are three types of fastening to explore.
The first option is a zip. At Klipspringer, we don’t typically recommend zips for SegriCovers as they can get stuck, are a less hygienic option, and tend to have a shorter lifespan than other fastenings.
The second option is Velcro. Although this method isn’t going to get stuck, it is similar to a zip in the sense that it is less hygienic and less likely to stand the test of time.
The third option, and our preferred approach here at Klipspringer, is magnets. This neat, hygienic, and durable fastening method is favoured by food production sites; making life easier for operatives who can simply pull open or snap together the relevant slits.
Measuring for a clear window
In addition to cut-outs, many food production sites will ask that their SegriCovers include clear panels. This is a useful customisation option if you need a screen to be visible even when your SegriCover is in place.
If you require a clear panel, it is essential that you list the size of the relevant screen, noting both the width and height. You will also need to measure where the screen sits in relation to the edge of the machine and the nearest corner.
Once you have supplied these measurements, we will be able to position the clear panel so that it sits directly above the screen in question.

Assessing your SegriCover quote

Once our team has received your measurements and come back to you with any relevant questions, we will be able to design the SegriCovers for your site.
We will then send through a quote that includes the dimensions of your SegriCovers.
To avoid any mistakes, it is essential you read this quote carefully and carry out a final check of the SegriCover dimensions.
At this stage, you may discover the dimensions of your SegriCover are slightly larger than the dimensions of your equipment. Typically an extra 10-20mm will be added to the width and depth of the SegriCover to ensure a tight fit, but one with enough give to allow your operatives to lift the cover on and off.
As for the height, the added length will often be an additional 50-100mm. This can help to weigh down the SegriCover and will prevent any liquids from splashing up inside. However, even with this extra length in mind, the SegriCover should never be touching the floor of your factory. This is because a cover that touches the floor is unhygienic and could become a point of non-conformity during an upcoming audit.
That brings us to the end of our guide to measuring your equipment for a SegriCover. We hope this article has provided a clear picture of the process involved and has given you all the information you need to confidently take your measurements.
If you have any remaining questions, our team of experts are on hand to help. You can contact them on 01473 461800Β orΒ sales@klipspringer.com. Alternatively, you can share your details using the contact form below.Β
If you want to learn more about SegriCovers or other segregation solutions, the Klipspringer team would be happy to help. Share your details below to arrange a consultation.
How to use a Comparator Pot to Check the Accuracy of your Infrared Thermometer
How to use a Comparator Pot to Check the Accuracy of your Infrared Thermometer
An important piece of equipment within the food production industry,Β infrared thermometersΒ indicate the amount of thermal radiation being emitted by the substance you are testing. Put simply, they allow you to find out the temperature of your product without having to touch it or penetrate it with a traditional thermometer probe.
As with any equipment at your site, infrared thermometers need to be suitably maintained, stored correctly, and treated with respect. It is also important that you verify your thermometers on a regular basis, and one of the most popular approaches involves the use of aΒ Comparator Pot.
NOTE: We want to be upfront and say that, at Klipspringer, we supply both Comparator Pots and an alternate method to verifying your thermometers β the LazaPort8. In this article we will be walking you through the process of using a Comparator Pot, but we will also be exploring the benefits of using the LazaPort8 instead.
With over 20 years experience in the industry and a dedication to providing honest and informative resources for our customers, we hope that you can trust our analysis of the two different approaches and understand the conclusion we arrive at.
How do I use a Comparator Pot?
Compact, lightweight, and easy to use, it’s no surprise that a lot of sites are using Comparator Pots to verify their infrared thermometers. These simple pots are especially popular at smaller factories with only a few thermometers in need of verification. If you are new to this method or would like a quick refresher, we’ve broken down the process into three simple stages:
#1 Prepare your Comparator Pot and Reference Probe

Take your comparator pot to the room where the verification will be taking place.
Leave it to stabilise for at least one hour or until it has acclimatised to the temperature of the room.
Near the base of the comparator pot, there will be a small hole for the penetration probe of your reference thermometer.
Insert the penetration probe into this hole, then leave your reference thermometer to stabilise.
#2 Use your Infrared Thermometer to Take a Reading

Take your infrared thermometer and hold it perpendicular over the top of the comparator pot.
Make sure that you are pointing the infrared thermometer at the black bottom of the comparator pot.
When you squeeze the trigger, use the laser guides on the front of the thermometer to position it correctly.
Squeezing the trigger will allow you to take a reading on your infrared thermometer.
#3 Compare the Readings of the Two Thermometers

We recommend taking three separate readings to ensure you get reliable results.
Compare the readings from your infrared thermometer to the readings of your reference probe.
Any difference will indicate the accuracy drift of the infrared thermometer. If it sits outside the parameters assigned by your site, your equipment will need to be sent off for external testing.
Record both readings on your calibration log.
What are the pros and cons of a Comparator Pot?
As with any piece of equipment, the Comparator Pot has positive and negative attributes. Before making an important investment, It’s vital that you evaluate these factors alongside the specific needs of your site. From cost and ease-of-use to accuracy and range of measurement, we’ve listed the key points to consider below:
- Time consuming process
- Less accurate results than the LazaPort8
- Unable to test a lot of thermometers at the same time
- Only allows verification at one temperature
- Unable to validate hot temperatures
- Although it is possible to place the pot in a freezer to drop the temperature to freezing, this is unreliable as the temperature will jump back up when the pot is returned to an ambient environment
- Small, simple device
- Budget friendly
- Suited to a small number of infrared thermometers
What is the alternative to a Comparator Pot?
As detailed above, the Comparator Pot does have a number of limitations. Consequently, a growing number of sites are switching to the LazaPort8. This piece of equipment will not only test your infrared thermometers, but will also verify your standard thermometers. Loved by auditors and customers for ultimate traceability, it couldn’t be easier to use:

The LazaPort8 has two temperature blocks on top – one at a low temperature such as -18 or 0Β°C and the other at 100Β°C.
Point your infrared thermometer at the top of one of the blocks and gain a stabilised reading.
We recommend taking three readings to ensure you achieve consistent results.
Repeat this process on the LazaPort’s other temperature block.
Record the results in your calibration log.
What are the pros and cons of a LazaPort8?
Listed below are the plus points and downfalls of a LazaPort8. Again, it is important that you take them into consideration before making a decision for your site. You can also contact our team onΒ 01473 461800Β orΒ sales@klipspringer.comΒ if you would like further information on this piece of equipment.Β
- Higher upfront cost
- Verifies multiple thermometers at once
- Tests the entire temperature range you will be testing your products at
- Highly accurate to +/- 0.3Β°C.
- UKAS calibration certificate proves conformance to the UKβs most common recognised standard
- Verifies probe and infrared thermometers
- Overall savings of money and time
What is the best approach to checking the accuracy of my thermometers?
Although we have taken you through the pros and cons of both the Comparator Pot and the LazaPort8, the final decision lies with you. After all, you are uniquely positioned to know what is best for your site. We do hope; however, this article has provided you with a balanced view of the two options available and helped you to understand our preference for the more advanced piece of equipment.
The video below should also be a valuable tool, providing helpful demonstrations of the Comparator Pot and LazaPort8 in use:
Both pieces of equipment are available to view and purchase online atΒ klipspringer.com. Alternatively, if you require further support with your decision, simply fill out the contact form below and one of our friendly team members will be in touch.
How to assemble your SegriScreen
How to assemble your SegriScreen, plus everything you need to know about cleaning, maintenance, and storage
SegriScreens have been designed for the specific purpose of segregation within the food production industry. Suited to everything from allergen segregation to washdown protection, they offer the following benefits:
- Mobile screens connect to form a continuous, connected barrier for washdown, production or maintenance use.
- Modular system - simply add or remove as many SegriScreens as required.
- Screens nest for convenient storage and transportation.
- Favoured by industry leads such as Bakkavor, Kerry, Morrisons, Greggs, and Greencore.
- Hygienically designed for food industry applications.
- Add bespoke print for personalised branding, safety or instruction messages.
Thanks to the signature design of the SegriScreen, they are also easy to assemble. In fact SegriScreen assembly should only require two operatives and around ten minutes of concentration.
Even so, we wanted to produce an online Help Guide that you can return to if you ever need to remind yourself of the process.
SegriScreen Assembly – Video Guide
Watch the video below for aΒ step-by-step tutorial on how to assemble your SegriScreen.
NOTE: This video shows you how to assemble a Master SegriScreen, which has two sets of wheels. Additional screens can then be added and they have just one set of wheels. Additional wheel sets can be purchased if required.Β
SegriScreen Assembly – Written Instructions
Alternatively, if you and your team would prefer to follow written instructions. You will find a step-by-step guide below:

1. Slot the bottom ends of the side bars into the wheel bases. The bases have side brackets that should face outwards on both sides.Β

2. Making sure the bolt holes line up, screw in two bolts with washers on each side. Tighten the fixings using a spanner.Β
NOTE: Master Screens (as shown in the video and images) have two wheel bases, one for each side. Additional Screens have one wheel base as the other sideΒ slots into the adjacent screen.Β

3. Take the bottom, horizontal bar and use two bolts to fix it between the corner brackets. The bolts should go through the long, oblong holes in the brackets.Β

4. Repeat this process with the top, horizontal bar, making sure the SegriScreen hooks are facing forward with their tips pointing down. Use bolts in each corner to hold in place.Β
NOTE: Only screw the bolts in part way, as this will allow you to tension the screen at a later stage.Β

5. Remove the packaging around the hooks on the top of the bar. Spread them along the bar until they are approximately equidistant from each other.Β

6. Wrap the SegriScreen material over the top of the frame, placing the hooks through the eyelets along the top. Hook the screen material along the side bars, one hook per eyelet.Β
NOTE: One of the sides has a larger flap than the other. This will allow it to overlap with any adjacent screens you wish to connect. Β Β

7. Pull the side bars outwards so the metal tabs at the end of the horizontal bar click into the slots in the top and bottom brackets. This will tension the screen.Β

8. Using a spanner, tighten up the bolts in all four corners to secure the frame. You can then attach Additional Screens to your Master Screen as required.Β
How to clean, maintain, and store your SegriScreens
When it comes to introducing any new equipment to your site, it is vital that you and your team are aware of the best approach to cleaning, storage, and maintenance.
This will help you to drive standards of hygiene and food safety. It will also help you to extend the lifespan of your equipment. With this in mind, we have put together a guide to effective SegriScreen maintenance.

NOTE: With the correct approach to cleaning, storing, and maintenance, your SegriScreens should stand the test of time. However, if an unexpected incident should occur, replacement screens are available – allowing you to replace the material whilst making use of the existing base.Β
To learn more about SegriScreens or the other segregation solutions available, you can navigate the buttons below.
You can also reach out to our support team on 01473 461800Β orΒ sales@klipspringer.com. Alternatively, you can share your details using the contact form below.Β
If you want to learn more about SegriScreens or other segregation solutions, the Klipspringer team would be happy to help. Share your details below to arrange a free consultation.
















Shadow Boards & Equipment Storage
Visual Management
Hygiene & Production Equipment
Food Handling & Storage Containers
Instrumentation and Monitoring
Calibration & Service
Segregation & Washdown Covers
Detectable Equipment
Specialist PPE
Stainless Steel Equipment
Hygiene Squeegees
Vacuum Cleaners
Food Manufacturing
Food Storage, Distribution & Retail
Auditors & Food Safety Inspectors
Food Service & Hospitality